SG-3013 Isolated RTD Input Signal Conditioner


Price:
Sale price£147.00

Description

The SG-3013 is a RTD input signal conditioning module to transform RTD input to voltage or current output. SG-3013 uses a microprocessor-based transducer, which integrating two high resolution ADCs and an MCU, to acquire the RTD signal. The supported RTD types are Pt100 , Pt1000 , Ni 120 ,Cu100 ,and Cu1000.

The SG-3013 supports 1000Vdc 3-way isolation. The power supply that drives the module's input and output circuitry is internally isolated, enabling SG-3013 to offer true channel - to - channel isolation.

It's easy to mount the SG-3013 on a standard DIN rail and it can operate in environments with wide temperature ranges.

Specifications

Analog Input

  • Input Channels: 1
  • Input Type: Pt100 α=0.00385 / Pt100 α=0.003916 / Ni 120 / Pt1000 α=0.00385
  • Input Connections: 2/3/4 Wires
  • Accuracy: ±0.1% of full range
  • Three-way isolation: 1000Vdc

Analog Output

  • Channels: 1
  • Voltage Output (Unipolar): 0~5V,0~10V (Output impedance : <50Ω)
  • Current Output: 0 ~ 20mA, 4 ~ 20mA
  • Current load resistor: 0~500Ω(Source)

Other Specifications

  • Weight: 94 gram
  • LED Display: 1 LED as Power/ Communication indicator
  • Power Input: +10 to +30 Vdc
  • Power Consumption: 1.2W

Environment

  • Operating temperature: -25°C to 75°C
  • Storage temperature: -30°C to 75°C
  • Humidity: 5 to 95%, non-condensing
Switch

Temperature Range (°C)

RTD Type Platinum 100,α= 0.00385
1

0x00

00000

-200~200

2

0x01

10000

-200~600

3

0x02

01000

-100~100

4

0x03

11000

-100~0

5

0x04

00100

-100~200

6

0x05

10100

-50~50

7

0x06

01100

-50~150

8

0x07

11100

0~100

9

0x08

00010

0~200

10

0x09

10010

0~600

RTD Type Platinum 100,α= 0.003916

11

0X0A

01010

-200~200

12

0X0B

11010

-200~600

13

0X0C

00110

-100~100

14

0X0D

10110

-100~0

15

0X0E

01110

-100~200

16

0X0F

11110

-50~50

17

0x10

00001

-50~150

18

0x11

10001

0~100

19

0x12

01001

0~200

20

0x13

11001

0~600

RTD Type Platinum 1000,α= 0.00385

21

0x14

00101

-200 ~ 600

RTD Type Nickel 120

22

0x15

10101

-80 ~ 100

23

0x16

01101

0 ~ 100

24

0x17

11101

0~300

25

0x18

00011

-80~300

RTD Type Cu 100 at 0°C,α= 0.00421

26

0x19

10011

-20 ~ 150

RTD Type Cu 100 at 25°C,α= 0.00427

27

0x1A

01011

0 ~ 200

RTD Type Cu 1000 at 0°C,α= 0.00421

28

0X1B

11011

-20 ~ 150

 

Technical

User Manual

Estimate shipping

You may also like